Controls

Hydraulics

Laser Particle Counter: Type LasPaC-II-P (Portable)

The LasPaC-II-P (Portable) is the most complete way to measure the contamination level of your system.
With the LasPaC-II-P you have the ability to measure, analyze and document your results immediately without the need of any additional equipment.

  • Specifications
    • + expand all
    • - collapse all

    Technical Data

    Dimensions (L/W/H) 551 x 358 x 226 mm / 21.69 x 14.09 x 8.90 in -
    Weight 13 kg / 28.66 lbs -
    Keyboard QWERTY keyboard -
    Printer Integrated thermal printer 384 dots per line
    Voltage Range 110 to 240 V AC / 12 to 24 DC European, UK and US power plug adaptors included
    - Number of tests before recharging is required: 100
    Calibration ISO Medium Test Dust (MTD) according to ISO 11 171:1999 -
    Analysis ISO 8-24, ISO 4406 Code, NAS 1638 Code 2-12, SAE AS 4059 Code 2-12 -
    Pressure Range 2 to 400 bar / 29 to 5,801 psi -
    Viscosity Range 1 to 400 cSt -
    Hose Connections Test coupling STAUFF Test 20 or comparable (M16 x 2) -
    Operating Temperature +5°C to +80°C / +41°F to +176°F -
    Data Output Cumulative particle counts, as well as cleanliness classes according to ISO 4406 (1999) / SAE AS 4059 Rev.D (2001) and ISO 4406 (1191) / NAS 1638 (1964) -
    Maximum Concentration ISO 24 -
    Accumulator Internal rechargeable battery -
    Data Storage 600 tests -
    Fluid Compatibility Mineral Oil, Petroleum based fluids Phosphate Ester and Water Glycol compatible devices on request
    Computer Interface RS-232 communication port as standard USB adaptors included
    External Alarm External alarm socket with switching outputs max. 24 V DC/AC, 1 A -
    Software Downloading and storage of the data with included “LasPaC-II View” software Further processing with Microsoft Excel® possible

    Laser Sensors

    High Accuracy Laser 4 to 6 μm(c) The orifice of the sensor has a cross section of 0.9 x 0.9 mm / .04 x .04 in
    Standard Accuracy Laser 6 to 68 μm(c) The maximum concentration is ISO 4406 Code 24 (160.000 p/ml)
    Measured Channels 4, 6, 14, 21, 25, 38, 50, 68 μm(c) -